List Price: On request
If you are not sure whether this unit is suitable for your application, please do not hesitate to contact us for further information
The detector is based on continuous thermal conductivity measurement of gases with MEMS sensors developed by Archigas.
*All data refer to 0.5 vol% H2 in N2
Specifications | |
Dimensions with connectors; weight | W=56mm, H=36mm, D=30mm; ~ 350g |
Power supply | 24 V DC (±6 V), 1 A |
Communication | RS-485 (19200 / 8bit) Optionally: USB / RS232 / Analog |
Ambient conditions | |
Ambient temperature range | -10°C to 50°C, depending on mounting and thermal insulation |
Ambient Humidity | 95% RH |
Gas analysis | |
Carrier gas | Any gases as H2 / He / Ar / N2 / O2 |
Pressure | 0.8 – 20 bar abs. |
Flow rate | 20 – 2500 ml /min |
Warm up time | Approx. 10 min |
T90-time | < 0.5 s |
Dynamic range | 50 – 106 ppm |
Noise | < 1 ppm* |
Drift at zero point | < 50 ppm per week |
Repeatability | < 50 ppm |
Error due to change of ambient temperature | < 50 ppm per 10°C |
Error due to change of flow at 70l/h | < 50 ppm per 10l/h |
Gas pressure (absolute) | 2000 kPa (20 bar) |
Error due to change of pressure (above 800 hPa) | < 50 % per 10 hPa |
High stability due to thermostatization of the system at 60 °C
For any further questions or quotes please do not hesitate to contact us using the contact form or contact information below.